SP25M user's guide
http://www.renishaw.com
Establishing a docking position for SM25 and TM25-20 modules
This section describes how to manually position the probe body and module to define the module docking position of the desired port. This
process requires very fine movements under CMM joystick control and care must be taken to avoid any collisions of the probe body, module
and port.
Eye protection should be work during this procedure and a good level of lighting is recommended to ensure no collisions occur. This
procedure is the same when using the FCR25 or the FCR25 TC rack.
1. Latch the port lids open using the triangular plastic inserts provided.
2. Place the SM25 or TM25-20 module in the desired port.
Steps 1 and 2 - latching port lids open and placing module into desired port (n)
3. Ensure the probe head is orientated at an A0 B0 position.
4. Mount the SP25M probe body to the probe head.
5. Inhibit the probe signal through software.
NOTE: Great care is required during the next part of this procedure as there is very little clearance between the probe body and
module and the rack port. Carefully observe all movement to ensure that no collisions occur.
6. Using CMM joystick control, position the probe head and SP25M probe body directly above the probe module that has been placed in
the rack port.
7. Slowly lower the probe body towards the probe module ensuring that the XY position of the probe body does not collide with the port.
Issued 10 2015
Step 6 onwards - watch for clearances to ensure no collision between parts