SP25M user's guide
http://www.renishaw.com
Put down routines
Put down routine - SM25 scanning modules and TM25-20 adaptor module
Move description
Clearance position
Enter port
Detach module
Clear of module
Exit port
To ensure optimum metrology performance it is recommended that following any module change the probe head is unlocked and then
relocked.
Put down routine - SH25 stylus holders
Move description
Clearance position
Enter port
Detach stylus holder(s)
Clear of module
Exit port
Put down routine - TP20 module
Move description
Clearance position
Enter port
Detach TP20 module
Clear of module(s)
Exit port
* Coordinates assume aligned with Y-axis as per
** Calculate nominal 'Z' docking position according to the stylus length used (see calculation formula in section
position for SH25 stylus
holders).
Issued 10 2015
X-axis (mm)*
[dat_MOD_port(n)]
no change
no change
no change
no change
X-axis (mm)*
[dat_SH_port(n)]
no change
no change
no change
no change
X-axis (mm)*
[dat_TP20_port(n)]
no change
no change
no change
no change
FCR25 mounting instructions
Y-axis (mm)*
[dat_MOD_port(n)] - 30 mm
[dat_MOD_port(n)]
no change
no change
[dat_MOD_port(n)] - 30 mm
Y-axis (mm)*
[dat_SH_port(n)] - 30 mm
[dat_SH_port(n)]
no change
no change
[dat_SH_port(n)] - 30 mm
Y-axis (mm)*
[dat_TP20_port(n)] - 30 mm
[dat_TP20_port(n)]
no change
no change
[dat_TP20_port(n)] - 30 mm
Z-axis (mm)*
[dat_MOD_port(n)]
no change
[dat_MOD_port(n)] + 3 mm
[dat_MOD_port(n)] + 8 mm
no change
Z-axis (mm)*
[dat_SH_port(n)]
no change
[dat_SH_port(n)] + 3 mm
[dat_SH_port(n)] + 8 mm
no change
Z-axis (mm)*
** [dat_TP20_port(n)]
no change
**[dat_MOD_port(n)] + 3 mm
** [dat_MOD_port(n)] + 8 mm
no change
Establishing the docking
42