Vorschau ausblenden Andere Handbücher für SP25M:
Inhaltsverzeichnis

Werbung

SP25M user's guide
http://www.renishaw.com
3. Ensure the probe head is orientated at an A0 B0 position.
4. Inhibit the probe signal through software.
5. Mount the desired SM25 scanning module to the SP25M probe body.
6. Mount the SHSP to the probe module as described above.
Step 6 - the SHSP is mounted to the SM25 scanning module
7. Enable the probe signal through software (probe signal now armed).
8. Turn off tip radius compensation.
9. Measure a point on the top of the port [P1], offset the point by -20 mm in the Z-axis.
Step 9 - take a point on port top surface
Step 10 - take a point on port front side face
Step 11 - taking 2 points on port inside rails
10. Measure a point on the front lip of the port [P2], offset the point by + 16 mm in the Y-axis.
11. Measure two points [P3 and P4] on the inside rails of the port and construct a point [P5] between them.
12. Create a datum coordinate system for the stylus holder docking position of the port. This will be at the X position of P5, the Y position of
P2 and the Z position of P1 – [dat_SH_port(n)].
Issued 10 2015