SP25M user's guide
http://www.renishaw.com
Step 6 and 7 - moving to approximate XY position then refining to ensure good alignment between body and module
8. Continue to lower the probe body towards the probe module until the magnetic attraction between the kinematic joint makes the module
move towards the probe body. The module should move slightly towards the kinematic joint of the probe body without tilting (indicating good
XY alignment). If any tilting occurs when the module moves towards the probe body (indicating poor XY alignment) then the probe head and
body should be repositioned and the process should be repeated.
Step 8 - slowly move body downward in Z axis until module jumps straight upwards due to magnetic attraction
9. Slowly lower the probe body on to the module, stop any movement when the LED on the probe head illuminates. This indicates that an
electrical connection is made.
Step 9 - slowly move body downward in Z axis - STOP immediately when the probe head LED is lit
10. Offset the CMM using a DCC movement of 0.75 mm in a –Z direction at a speed of 5 mm/s.
11. Create a datum coordinate system for the module docking position of the port at the following position: [dat_MOD_port(n)].
12. Slowly move the connected probe head, body and module assembly in a –Y direction clear of the port.
13. Remove the module and place back in the port.
14. Enable the probe signal through software (probe signal now armed).
15. Repeat steps 2 to 13 above for all other modules required.