Description of Device:

The ON-Q* Pump with Fixed Flow rate delivers medication at
a continuous flow rate.
1
E-Clip (100 ml vol. or less)
2
Fill Port Cap
3
Fill Port
4
ON-Q* Pump
5
Clamp
6
Air-eliminating Filter
7
Tubing
8
Flow Controller (tape to skin)
Figure 1
8
(Figure 1)
1
2
3
4
5
6
7
Medication Label:
(Figure 2)
Medication Label Symbol Definitions:
1
Patient Name
2
Date
3
Time
Figure 2
1
2
3
Instructions for Use
Use Aseptic Technique
Filling the ON-Q* Pump: (Figure 3)
Follow hospital protocols and applicable regulations for
Note:
filling pump.
1. Close clamp.
2. Uncap the fill port.
3. Attach filled syringe to fill port.
4. Invert pump as shown.
5. Grasp syringe with both hands.
6. Push down on plunger continuously until volume is
dispensed. Do not handle pump while filling, as the
syringe tip may break. Repeat as necessary. (Syringe
accuracy is ±4%).
Note: Filling Extension Set is provided with larger pumps
(see product insert for further information).
Caution: Do not underfill pump. Underfilling the pump
may significantly increase the flow rate. Do not
exceed maximum fill volume. (Table 1)
7. Remove filling device from fill port.
8. Replace fill port cap. Label with the appropriate
pharmaceutical and patient information.
Note: The ON-Q* Pump contains either an E-Clip or
Carry Case for holding the pump.
4
Medication
5
Dosage
4
5
5