Herunterladen Inhalt Inhalt Diese Seite drucken

Activating And Deactivating A Display; Low Resolution Function - WIKA Obsolete CPH65I0 Betriebsanleitung

Eigensicherer hand-held druckkalibrator
Vorschau ausblenden Andere Handbücher für Obsolete CPH65I0:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6 Commissioning, operation

6.5.5 Activating and deactivating a display

Use the "DISPLAY" option on the 'Display selection' main menu to access the display activation
menu.
GB
The F2 button can be used to select the desired display. The "ON/OFF" option switches the active
display on or off. The active display and the current status "ON/OFF" are shown in the lower
display.
Select the "DONE" option to save the changes and return home. When a display is deactivated its
configuration is retained. As soon as the display is activated, the configuration is checked against
the configurations of the other currently-active displays. If there is any conflict between
configurations, the recalled display's configuration is modified to avoid the conflict. If all three
displays are deactivated the lower display will switch on automatically.

6.5.6 Low resolution function

Due to the high accuracy of the CPH65I0 the measured values are displayed with many digits, this
might be a disadvantage in some cases. Therefore the CPH65I0 has a low resolution function. This
function takes away the last digit.
To turn the function on or off, proceed as follows:
With the calibrator turned on and operating press the F3 key to activate the "MORE" menu option.
Press the "NEXT" button until "RESOLUTION" appears in the left text field. Now press the F1 key
to enter the function.
Select ON or OFF to turn the low resolution function on or off.
Use "DONE" to return to main menu.
WIKA Operating Instruction, Model CPH65I0
23

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis