6 Commissioning, operation
is chosen the configuration is unlocked and the menu display continues to the next sub-main
menu.
GB

6.5.3 Saving and recalling settings

The calibrator will automatically save the current set-up for recall at power-up. Additionally 5 setups
can be accessed through the "SETUPS" menu. Select the "SETUPS" option from the "Setups
main" menu.
"SAVE" to save the setting, "RECALL" to recall the setting or "DONE" to return home.
If "SAVE" or "RECALL" is selected, use the arrow keys to select the setting location. Then use the
"SAVE" option to store the current setting into the selected location or the "RECALL" option to
recall the setting stored in the selected location. The display menu will automatically return home.

6.5.4 Setting auto shut-off parameter

The calibrator will switch itself off after a specified time; this function can also be deactivated. Use
the "AUTO OFF" option to set the auto shut-off parameters.
Use the F2 and F3 keys to select the time after which the calibrator will turn itself off, or disable the
auto shut-off by scrolling down to "0".
Use the "AUTO OFF DONE" option to set the parameters and return home. The auto shut-off time
is reset whenever a key is pressed.
22
WIKA Operating Instruction, Model CPH65I0