7 Maintenance, cleaning and servicing
7.3

Cleaning

CAUTION!
Before cleaning, correctly disconnect the CPH65I0 from the pressure supply, and
switch it off.
Clean the CPH65I0 with a soft cloth dampened with water.
Wash or clean the dismounted CPH65I0 before returning it, in order to protect
staff and the environment from exposure to residual media.
Residual media in dismounted CPH65I0 can result in a risk to persons, the
environment and equipment. Take sufficient precautionary measures.
To avoid personal injury or damage to the calibrator, do not allow water into the
case.
To avoid damaging the plastic lens and case, do not use solvents or abrasive
cleansers.
For information on returning the CPH65I0 intrinsically safe hand-held pressure calibrator see
chapter "8.2 Return".
7.4

Recalibration

DKD certificate - Certificates:
We recommend that the instrument is regularly recalibrated by the manufacturer, with time
intervals of approx. 12 months. Every factory recalibration includes, additionally, an extensive free-
of-charge check of all system parameters with respect to their compliance with the specification.
The basic settings will be corrected if necessary.
WIKA Operating Instruction, Model CPH65I0
GB
37