6 Commissioning, operation
6) Use the arrow keys to set lower limit of the measuring range, and select "DONE SET" when
finished. The "%-ERROR" mode will be ready to use.
The lower and upper limit of the measuring range will be saved in non-volatile
memory until they are changed again by the user for the internal sensors.

6.11 "MIN/MAX" value capture

The CPH65I0 pressure calibrators have a "MIN/MAX" function for capturing the minimum and
maximum values of any displayed parameter.
The "MIN/MAX" function can be set by stepping through the menu options until "MIN/ MAX" is
shown on the display above the F1 key. At this point, pressing the F1 key will scroll the display
through the "MIN/MAX" values that are stored in the "MIN/MAX" registers. These readings are live
so new "MIN/MAX" values will be recorded while in this mode.
To clear the "MIN/MAX" registers, simply press the "CLEAR" key. These registers are also cleared
on power-up or when the configuration is changed.
WIKA Operating Instruction, Model CPH65I0
GB
33