Problem:
Device does not
operate at all
No coolant in the
instrument
[ HOT ] appears in
display
Foot control fails to
function
In the event a problem cannot be solved, contact dealer or authorized service technicians, as listed on
the last page of your Operating Instruction Manual.
Accesories
Clip Set motor cable (10 pcs.) ............................................................................................................1873
Clip Set contra angle (3 pcs.) .............................................................................................................1881
Tubing (16cm) for Y-Connector ..........................................................................................................1773
Disposable tubing set, standard, sterile..............................................................................................1706
Y-Connetor .........................................................................................................................................1777
Cooling fluid flask, 0.9% NaCl, 1l........................................................................................................1707
Nou-Clean; Cleaning spray.................................................................................................................1984
Spray nozzle attachment E-coupling, for surgery instruments ...........................................................1958
Spray nozzle attachment, for electronicmotor ....................................................................................1942
For ordering other parts, our customer service personnel are pleased to assist
Disposal of device, components and accessories must strictly conform to local laws and regulations as
set out by the relevant authorities.
With regard to the preservation of the environment old equipment may be returned to the distributor or
manufacturer.
10

7 Troubleshooting

Cause:
Control unit not turned on
Wrong operating voltage
Mains not connected
Irrigation pump not turned
ON
Tubing set not mounted
properly
Tubing set is glued
together/coated with
deposit
Cooling fluid flask
not ventilated
Tubing set is dripping
Electronicmotor is getting
too hot
Foot control not connected

8 Replacement parts and numbers

9 Disposal

Solution:
Turn main switch
"ON/OFF"ON
Check mains voltage
output
Connect mains to control
unit
Turn the irrigation pump
ON
Mount tubing properly (pay
attention to direction!)
Exchange tubing
Open air filter on drip
chamber
Exchange tubing set
Turn off the unit, wait for
10 minutes, turn on the
unit again
Plug foot control into
control unit
NOUVAG AG
Refer:operating
Manual:
System Assembly
3
Technical Data
1.2
System Assembly
3
Operation with vario-
5.5
pedal
Set-up
4.1
Operation
4
System Assembly
3
Set-up
4.1
Set-up
4.1