Herunterladen Inhalt Inhalt Diese Seite drucken

Mikromotor Endo; Troubleshooting; Disposal - Nouvag TCM Endo III Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

NOUVAG AG

6.2 Mikromotor Endo

1)
The micromotor, cable and plug should be cleaned after each use to prevent the build up of
deposits and debris that can destroy it, if not removed. Wipe it with a clean cloth dampened with
disinfectant solution.
2)
Put the sterilizable stopple on the motor.
3)
After packing motor and cable, autoclave it at maximum 134° C .
4)
After autoclaving, allow motor to dry for one hour at room temperature.
Store motor bagged and tagged with sterilization date!
Do not bend the motor cable!
Clean, disinfect and sterilize the Micromotor after each use
Do not clean Micromotor with compressed air
Problem:
Device does not operate at
all:
Footswitch fails to function:
Micromotor does not run:
In the event a problem cannot be solved, contact dealer or authorized service technicians, as listed on
the last page of your Operating Instruction Manual.
Disposal of device, components and accessories must strictly conform to local laws and regulations as
set out by the relevant authorities.
With regard to the preservation of the environment old equipment may be returned to the distributor or
manufacturer.
No. 31817 29/05

7 Troubleshooting

Cause:
Plug is not inserted properly
Wrong operating voltage
Pedal not connected
Control unit not turned on
Motor not turned on
Motor not connected
Motor incorrectly attached

8 Disposal

TCM Endo III
Solution:
Insert plug and check fitting
Check mains voltage output
→ please see operating voltage on the
rating plate
Plug pedal cable into rear of control
unit
Turn main switch ON/OFF
Turn motor with „Motor"-key on or with
footswitch
Connect motor cable to control unit
Press handpiece firm on motor until
handpiece snaps in
7

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis