Intended Use of the Measuring
Instrument VC 240:
Measurement of direct voltages up to max. 1000VDC in the over-
voltage category CAT II
Measurement of alternating voltages up to max. 750VACrms in the
over-voltage category CAT II
Measurement of direct and alternating currents up to max. 20A, for
a maximum period of 10s with pauses of 15min between the single
measurements (secured)
Measurement of resistance values up to max. 20 MΩ
Continuity test (below approx. 50 Ω acoustic) and diode test
Measurements must not be carried out under unfavourable ambient
conditions. Unfavourable ambient conditions are the following ones:
- wetness or excessive humidity
- dust or combustible gases, vapours or solvents
- electrical storms or stormy conditions and strong electrostatic
fields, etc.
Use other than that described above will lead to damage to the
measuring instrument and additional risks, such as short circuit, fire,
electric shock etc. No part of the product may be changed or modi-
fied! The safety instructions should be observed without fail!
Operating Elements
Illustration (foldout page)
1a 20A input
This measuring input is secured by 10A
(fuse) and has been approved for direct
and alternating currents of up to max. 20A
(for a max. period of 10s with pauses of
15min between the single measurements).
22
Courant continu DCA (µ A et mA et A)
200 µA
±(1,0%+5 ch) 0,1 µA
2 mA
±(1,0%+5 ch) 0,001 mA
20 mA
±(1,0%+5 ch) 0,01 mA
200 mA
±(1,0%+5 ch) 0,1 mA
20 A
±(1,5%+5 ch) 0,01 A
Courant alternatif ACA (µ A et mA et A)
200 µA
±(1,2%+5 ch) 0,1 µA 40-100 Hz/100-400Hz ±(1,5%+10dgts)
2 mA
±(1,2%+5 ch) 0,001 mA40-100 Hz/100-400Hz ±(1,5%+10dgts)
20 mA
±(1,2%+5 ch) 0,01 mA 40-100 Hz/100-400Hz ±(1,5%+10dgts)
200 mA
±(1,2%+5 ch) 0,1 mA 40-100 Hz/100-400Hz ±(1,5%+10 ch)
20 A
±(2,0%+5 ch) 0,01 A 40-100 Hz/100-400Hz ±(2,0%+10 ch)
Résistance
200 Ω
±(1,0%+3 ch)
2 kΩ
±(1,0%+3 ch) 0,001 kΩ
20 kΩ
±(1,0%+3 ch) 0,01 kΩ
200 kΩ
±(1,0%+3 ch) 0,1 kΩ
2 MΩ
±(1,0%+3 ch) 0,001 MΩ
20 MΩ
±(1,2%+5 ch) 0,01 MΩ
Contrôleur de continuité
Il y a signal acoustique d'une continuité en cas de résistances
inférieures à env. 50 Ω.
Essai de diodes
La tensionde mesure maxi. sur un circuit de mesure ouvert est
d'env. 3 VDC maxi., résolution d'affichage 1 mV
Grandeurs d'entrée maximum, protection de surcharge
Mesure de la tension
: 1000 VDC ou 750 VACrms
Mesure du courant
: 20 A CA/CC dans la plage A 10 secon-
des maxi.,
avec ensuite une phase de refroidisse-
ment d'au moins 15 min.
0,1 Ω