3. Select ProbeCh. in the bottom menu, tips about probe checking shows on the
screen.
4. Select ProbeCh. again to begin probe checking and the checking result will occur
after 3s; push any other key to quit.
● Pass/Fail
The Pass/Fail function monitors changes of signals and output pass or fail signals by
comparing the input signal that is within the pre-defined mask.
Push the Utility button, select Function in the bottom menu, select Pass/fail in the
left menu.
The description of Pass/fail Menu is shown as the follows:
Function Menu
operate
Output
Horizontal
Rule
SaveRule
Pass/Fail test:
Detect whether the input signal is within the limits of the rule, if it exceeds limits of
the rule, it is "Fail"; otherwise it is "Pass". Also it can output fail or pass signal by
built-in and configurable output port. To run the test, read the following steps:
1. Push the Utility button, select Function in the bottom menu, select Pass/fail in the
left menu.
2. Enable switch on: Select Operate in the bottom menu, select Enable in the right
menu as ON.
3. Create rule: Select Rule in the bottom menu. Select Source in the right menu,
select the source in the left menu. Set Horizontal tolerance and Vertical tolerance
in the right menu. Select Create in the right menu to create the rule.
4. Set output type: Select Output in the bottom menu to enter output option setting.
Choose any one or two of the options "Pass", "Fail" or "Beep". "Pass" and "Fail"
are mutually exclusive options, which could not be chosen simultaneously. "Stop"
4.Advanced User Guidebook
Setting
Description
Enable
Control enable switch
Operate
Control operate switch
Pass
Signal tested corresponds with the rule
Fail
Signal tested not correspond with the rule
Beep
Beep when it satisfies the rule
Stop
Stop once satisfying the rule
Info
Control the display status of info frame
Source
Select source CH1, CH2 or Math
Change the Horizontal tolerance value by turning the
M knob
Vertical
Change the Vertical tolerance value by turning the M
knob
Create
Use the rule set as testing rule
Number
Select any one from Rule1 - Rule8 as your rule name
Save
Select Save to save the rule
Load
Load some rule as the testing rule