windows, trigger source is CH1, polarity is positive, 0.00mV the differential between
up level and low level threshold.
Windows Trigger menu list:
MENU
SETTING
Single
Windows
Mode
CH1
Source
CH2
Up Level
Threshold
Low Level
Polarity
Condition
Auto
Normal
Single
Mode
Holdoff
Holdoff
Reset
4.Advanced User Guidebook
INSTRUCTION
Set vertical channel trigger type as Windows trigger.
Select CH1 as the trigger source.
Select CH2 as the trigger source.
Adjust the M knob or click
threshold.
Adjust the M knob or click
threshold.
Positive Polarity, the oscilloscope triggers on the
positive Windows pulse.
Positive Polarity, the oscilloscope triggers on the
negative Windows pulse.
Enter: Triggers when the trigger signal enters the
specified trigger level range.
Exit: Triggers when the trigger signal exits the
specified trigger level range.
Time: Specify the hold time of the input signal after
entering the specified trigger level. The oscilloscope
triggers when the accumulated hold time is greater than
the windows time. Available range is 30ns-10s, default
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then stop
100 ns - 10 s, adjust M knob or click
interval before another trigger occur, press
panel button or click
which digit to be set.
Set Holdoff time as 100 ns
40
to set the up level
to set the low level
100ns.
to set time
to move cursor to choose