5. Commissioning and operation
Setting the sensitivity
The rotary switch on the right-hand side allows the sensitivity of the vibrating level switch
to be set. The sensitivity is preset to level 3 at the factory. If the medium to be measured
has a higher density than water, the vibrating level switch is triggered earlier (below the
switch point marking). By setting the sensitivity level to a value higher than level 3, this
EN
effect can be counteracted. If the medium to be measured has a lower density than
water, the vibrating level switch is triggered later (above the switch point marking). By
setting the sensitivity level to a value lower than level 3, this effect can be counteracted.
Rotary switch position 0:
Rotary switch position 9:
Switching delay
Delay setting
0
1
2
3
4
6
8
10
15
20
Housing cover interlock
Max/Min switch button
The Max-Min switch button is located on the right-hand side. Press this button to switch
between the maximum level and the minimum level.
5.7 Status LED
The following tuning fork types have a status LED to indicate the switching and error
states.
TLS-S
TLS-C/H with coupler connector in accordance with DIN 175301
TLS-C/H with M12 coupler connector
22
Highest sensor sensitivity
Lowest sensor sensitivity
Dial
0
1
2
3
4
5
6
7
8
9
WIKA operating instructions, models TLS-S, TLS-C, TLS-H
Max/Min
Indicator light
switch button
Sensitivity controller