5. Commissioning and operation
5.2 Switch point
The switch point of the tuning fork is marked by a horizontal notch when installed
vertically. The switch point mark should be at the desired height of the switch point
during installation. This switch point is set at the factory using "water" as a test medium.
If the density of the medium deviates from that of the test medium of water, the switch
point of the instrument shifts and can be adjusted via the sensitivity. The sensitivity can
only be set for the model TLS-S. The models TLS-C and TLS-H tuning forks do not offer
the option of adjusting the sensitivity.
The tuning fork body is available in two different lengths: 38 mm [1.50 in] or 42 mm
[1.65 in].
Tuning fork body, 38 mm [1.50 in]
Mounting situation
Tuning fork mounted vertically downwards
Tuning fork mounted vertically upwards
Tuning fork mounted horizontally
Tuning fork length, 38 mm [1.50 in]
WIKA operating instructions, models TLS-S, TLS-C, TLS-H
Switch point
13 mm [0.51 in], measured from the lower
edge of the vibrating body tip
32 mm [1.3 in], measured from the upper edge
of the fork extension
4 mm [0.16 in] above the centre of the tuning
fork