6.2. Measuring DC/AC Current
WARNING!
*
Do not apply voltage directly across the terminals. You
must connect the meter in series with the circuit.
*
The 10A terminal is fused. A severe fire hazard and short
circuit danger exists if you apply a voltage with high-
current capability parallel to this terminal. The meter can
be destroyed under such conditions.
To measure current, break the circuit and connect the probes
to two circuit connection points in series. Never connect the
probes across a voltage source in parallel. Doing so can blow
the fuse or damage the circuit under test.
Note: The maximum input current is 0,2 A or 10A depending
on the terminal used. In the 10A range excessive current flow
blows up the fuse, which must be replaced.
1. Set the rotary selector to the desired A range.
2. Use SELECT key to switch between AC or DC
measurement
3. Plug the black test lead into your meter's COM terminal
and the red test lead into your meter's mA or 10A
terminal.
4. Connect the test leads in series with the circuit and read
the measurement value.
51