7.2. Replacing the fuse
WARNING! To avoid electric shock disconnect the test leads
before removing the battery or the fuse. Replace only with the
same type of battery or fuse. Service should be performed
only by qualified personnel.
Caution! For continued protection against fire or other
hazard, replace only with a fuse of the specified voltage and
current ratings.
F1 200mA / 1000V (FF): 6,3 x 32 mm (Model: P 7640)
Breaking Capacity: 10 kA
F2 10A / 1000V F: 10 x 38 mm (Model: P 7530)
Breaking Capacity: 50 kA (AC), 30 kA (DC)
Follow these steps to replace the fuse:
1. Turn off the meter and disconnect the test leads.
2. Remove the screw on the battery cover, remove the
battery cover and remove the old battery, too.
3. Remove the defective fuse and replace it with a new
one with the correct specified voltage and current
ratings.
4. Replace the battery and replace the battery cover
on the case and secure it with the screw.