When taking measurements make sure no objects are trapped between the
clamp jaws (e.g., cables).
Do not exceed the maximum permissible measurement values.
Risk of fatal electric shock! The product must never be used if the housing or
battery compartment cover is open.
Risk of electric shock! Use caution when working with voltages above 30 V/AC
rms (42.4 V peak), 60 V/DC.
The rotary switch should be set to the correct range / function before each use.
Inspect the product for damage before each measurement. Never take meas-
urements if the insulation or product is damaged.
Use extreme caution when working around bare conductors or bus bars as
contact can result in electric shock.

6.8 Test leads and probes

The voltage between the connection points of the meter and the earth potential
must not exceed 300 V AC/DC in CAT III.
Probe assemblies to be used for MAINS measurements should meet EN
61010-031 standard, rated CAT III 300 V, 10 A or better.
Risk of electric shock! Use caution when working with voltages above 30 V/AC
rms (42.4 V peak), 60 V/DC.
Test leads must be disconnected before changing the range / function.
The cables have a wear indicator. If damaged, a second insulation layer in a
different colour becomes visible. Do not use if this occurs, replace immediately!
When taking measurements do not grip beyond the finger barrier or grip range
markings on the probes.
When taking CAT III measurements, probes
with cover caps (max. 4 mm free contact
length) must be used to avoid accidental
short circuits.