CAUTION
Danger of tilting or damaging cart and equipment.
Do not use the handles to lift the cart, only to push it
Only move the cart in a controlled manner and at walking pace.
Watch out for obstacles. Do not bump against walls or doors.
Take special care when negotiating inclinations, thresholds or uneven ground (e.g. in an
elevator). Always secure loose small parts.
Secure or if necessary remove and separately transport sensitive and heavy equipment
(e.g. monitors).

18 Warranty and Repair

The manufacturer provides a warranty of twelve months as from the delivery date on the chassis, converter,
charger and entire cabling of the mobile workstation. Excluded from the warranty is damage due to natural
deterioration (batteries), overloading or incorrect use/handling. If any modification is made to the mobile
workstation without consulting the manufacturer, this completely extinguishes the warranty.
If malfunctions of the mobile workstation occur, it must be returned after consultation to the manufacturer. In
this event, make sure you have the serial number, date of purchase and a precise description of the
malfunction ready.

19 Maintenance

19.1 General information

The mobile workstation is maintenance-free. Check the rollers and the brakes every year.

19.2 Socket strip with residual-current circuit breaker (RCD)

NOTE
The socket strip with RCD is only installed in mobile workstations with DC/AC Inverter.
The 2-pin socket strip with grounding contact is equipped with a residual-current circuit breaker (RCD).
NOTE
The RCD test button must be pressed regularly by a specialist electrician or a person briefed in electrical
engineering. The test deadlines are defined in the DGUV directive 3 and DGUV directive 4 according to
§5 paragraph 1 (Reference value: Every 6 months).
The RCD must be tested when the mobile workstation is switched on.