Trevira CS (fabrics)
Removable covers made of Trevira CS can be dry
cleaned� Washing under easy-care conditions at 60°C
is possible� Do not dry and do not iron!
Dartex
Test washed by the manufacturer at temperatures be-
tween 40 and 95 °C�
Accessory components made of metal, plastic, wood
Please clean accessory components with a conven-
tional mild household soap solution�

11.1.3. Disinfection of the product

Please use a water-based agent for disinfection of
the product, e�g� Terralin, Quartamon Med or Sagro-
tan Original concentrate� Please observe the manu-
facturer's instructions for use when doing so� When
disinfecting the fabrics, also follow the instructions
of the fabric manufacturer regarding the use of brand
name disinfectants�
Rinse off all detergents and disinfectants thoroughly
and leave the cleaned object to dry completely before
use� If this is neglected, this can lead to chemical reac-
tions making the material brittle or lead to skin irrita-
tions of the patient�
Important!
No disinfectants for synthetic leather/skai
Synthetic leather covers/skai covers
may not be cleaned with disinfecting agents, as
the material gets porous over time and cracks�
Note!
Wipe disinfection
If wipe disinfection is necessary, use a
disinfectant that is suitable for surface
disinfection according to the DGH disinfectant list�
Checking / components
Visual inspection of the seating system and the attached components and
accessories for visible damage
Check of connection between undercarriage and seating system
Visual inspection of all fastening elements
Visual inspection of all accessories / options
Retightening of all fastening elements
Retightening of all screw connections of the accessories / options
11.3.

Service

Test / Execution
Cleanliness
Check of all settings of the MAGICLIGHT
Check of setting of the head rest:
Does the head rest lie against the patient's occiput (standard version)?
Does the bottom edge of the head rest lie against the base of the head (shell shape)?
Do the side elements exert pressure on the patient's head?
Does the patient have sufficient freedom of movement?
Check of setting of the foot rest:
Is there enough free space at the back of the knee?
44
MAINTENANCE PL AN

11.2. Maintenance

Your interco product is checked by trained and qualified
employees for proper functioning and freedom from
defects before being handed over to the user� To ensure
safe use of the product over its entire service life (ac-
cording to the manufacturer from at least 2 years to
6 years maximum) and to ensure a perfect technical
condition, it must be maintained and inspected regu-
larly according to the following maintenance (every 6
months) and service (every 12 months) plans�
Note!
Changing settings
The changing of settings (other than
those described in Section 7) is only
permitted for qualified and trained personnel!
Incorrect settings by the user, operator or laypersons
bear the risk of injuries or crushing, which in a worst
case scenario could lead to the patient's death!
During normal use, you should regularly check the
interco product for damage and the perfect working
condition of all components�
Important!
Original spare parts
Only original parts may be used for
service and maintenance work�
seating system
®
Daily
Weekly
x
x
x
x
x
x
OK
Adjustments
necessary