Herunterladen Diese Seite drucken

Seca 655 Bedienungsanleitung Seite 191

Vorschau ausblenden Andere Handbücher für 655:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Fault/error code
No beeps audible
Measured results are not an-
nounced
080-297XX-XXX
081-297XX-XXX
082-297XX-XXX
083-297XX-XXX
Cause
Function not activated
Volume set to zero
Loudspeaker is defective
Function not activated
Volume set to zero
Loudspeaker is defective
Voice output memory cannot be read
Voice file not found
An error occurred during the measure-
ment
An error occurred during calibration
Remedy
è Activating/deactivating
Activate function
beeps, page 181
è Setting the vol-
Increase volume
ume, page 181
Inform seca Service
è Activating/deactivating
Activate function
announcement of measured results (Re-
sults), page 183
è Setting the vol-
Increase volume
ume, page 181
Inform seca Service
Inform seca Service
Deactivate beeps and voice output to sup-
press the error message until repair is car-
ried out:
• Deactivate announcement of patient in-
è Activating/deactivating an-
structions
nouncement of patient instructions (Mea-
surement), page 182
• Deactivate announcement of measured
è Activating/deactivating an-
results
nouncement of measured results (Re-
sults), page 183
è Activating/deactivat-
• Deactivate beeps
ing beeps, page 181
Inform seca Service
Deactivate beeps and voice output to sup-
press the error message until repair is car-
ried out:
• Deactivate announcement of patient in-
è Activating/deactivating an-
structions
nouncement of patient instructions (Mea-
surement), page 182
• Deactivate announcement of measured
è Activating/deactivating an-
results
nouncement of measured results (Re-
sults), page 183
è Activating/deactivat-
• Deactivate beeps
ing beeps, page 181
Repeat measurement and ask the patient
to keep still
If the error recurs, inform seca Service
• Remove objects from the immediate
vicinity of the device.
Troubleshooting • 191
English

Werbung

loading

Diese Anleitung auch für:

654455257256254453