Seca 655 Bedienungsanleitung Seite 137

Vorschau ausblenden Andere Handbücher für 655:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

5.3
Setting up device
To achieve accurate measured results, the floor at the setup location must be
level and stable. Soft floors (wooden boards, for example) give under the pa-
tient's weight and falsify the measured result.
1. Place the device on a firm, level surface.
2. Only for devices with an ultrasonic measuring rod: Mark the area shown in
the illustration using colored adhesive tape, for example.
min
min. 0,5 m
min
Example device configuration: seca 655 scale
seca 455 handrail, seca 257 ultrasonic measuring rod
NOTICE!
Malfunction caused by other ultrasonic emitters
If there are other ultrasonic emitters in the immediate vicinity of the de-
vice - automatic door openers, for example - incorrect measurements
will result.
► Ensure that there are no other ultrasonic emitters in the same room
or in the immediate vicinity of the device.
NOTICE!
Faulty measurement due to reflections
If there are objects or people in the immediate vicinity of the device, in-
correct measurements will result.
► Ensure that there are no objects or people within 0.5 meters of the
front or side of the scale during the measurement procedure.
► Ensure that the device is at least 0.2 meters away from the wall.
► Ensure that the patient is not wearing any kind of hair accessory on
top of their head.
min. 0,2 m
min. 0,2 m
min. 0,2 m
min. 0,2 m
,2
,2
m
m
min. 0,5 m
min. 0,5 m
min. 0,5 m
Starting up device • 137