Handling batteries and rechargeable batteries
120 • seca 655/654; seca 455; seca 257/256/254; seca 453
NOTICE!
Inconsistent measuring results
► Before you save and continue using measured values determined
using this device (e.g. in seca software or in an EMR system), en-
sure that the measured values are plausible.
► If measured values have been sent to seca software or to an EMR
system, ensure before continuing to use them that the measured
values are plausible and assigned to the correct patient.
NOTICE!
Malfunction caused by other ultrasonic emitters
If there are other ultrasonic emitters in the immediate vicinity of the de-
vice - automatic door openers, for example - incorrect measurements
will result.
► Ensure that there are no other ultrasonic emitters in the same room
or in the immediate vicinity of the device.
NOTICE!
Faulty measurement due to reflections
If there are objects or people in the immediate vicinity of the device, in-
correct measurements will result.
► Ensure that there are no objects or people within 0.5 meters of the
front or side of the scale during the measurement procedure.
► Ensure that the device is at least 0.2 meters away from the wall.
► Ensure that the patient is not wearing any kind of hair accessory on
top of their head.
WARNING!
Injury
Batteries and rechargeable batteries contain harmful substances
which may explode and so escape if not handled properly.
► Do not try to recharge batteries.
► Do not expose batteries/rechargeable batteries to heat.
► Do not burn batteries/rechargeable batteries.
► If acid is leaking out, avoid contact with the skin, eyes, and mucous
membranes. Rinse affected areas of the body with plenty of clean
water and seek medical help at once.
NOTICE!
Damage to device and malfunction due to incorrect handling
► Use only the type of battery/rechargeable battery quoted in this
document.
► Always replace all batteries/rechargeable batteries at the same
time.
► Do not short-circuit batteries/rechargeable batteries.
► If you do not use the device for an extended period of time, remove
the batteries/rechargeable batteries. This prevents acid from leak-
ing into the device.
► If acid has penetrated the device, do not continue using it. Have
the device checked by an authorized seca service partner and re-
paired if necessary.