Fault/error code
013-272XX-XXX
016-272XX-XXX
019-272XX-XXX
020-272XX-XXX
to
023-272XX-XXX
024-272XX-XXX
10.3
Ultrasonic height measurement
Fault/error code
symbol appears during
calibration
Ultrasound measuring head
power LED does not light up
Foot silhouettes on the
weighing platform do not illu-
minate
Patient instructions are not
announced
190 • seca 655/654; seca 455; seca 257/256/254; seca 453
Cause
Scale was caused to oscillate and was
unable to determine the zero point
Maximum capacity exceeded
Ambient temperature too high or too
low
One corner of the scale has been
loaded excessively
No GAL value
Cause
Calibration failed
Device is in an undefined state follow-
ing implausible input
Wiring in the ultrasound measuring
head incorrect
Power LED is defective
Device is in an undefined state follow-
ing implausible input
Foot silhouette illumination is defective
Function not activated
Volume set to zero
Loudspeaker is defective
Remedy
è Restarting the device, page 196
• If the error recurs, inform seca Service
Ask patient to step off the scale
• Observe ambient conditions for opera-
tion, transport, and storage
technical data, page 205
• Distribute weight evenly
è Restarting the device, page 196
• If the error recurs, inform seca Service
• Inform seca Service
Inform seca Service
Remedy
• Ensure that no objects or people are in
the immediate vicinity of the device during
calibration
• Ensure that the supplied calibration rod
was used
• Ensure that the calibration rod is posi-
tioned centrally on the foot silhouettes of
the weighing platform
è Restarting the device, page 196
• If the error recurs, inform seca Service
Wire the ultrasound measuring head as de-
scribed in the corresponding installation in-
structions
Inform seca Service
è Restarting the device, page 196
• If the error recurs, inform seca Service
Inform seca Service
è Activating/deactivating
Activate function
announcement of patient instructions
(Measurement), page 182
è Setting the vol-
Increase volume
ume, page 181
Inform seca Service
è General