Selecting BMI/BSA/Waist circumf.
8. Place the calibration rod centrally on the illuminated foot silhouettes of the
weighing platform.
9. Step back from the measuring device (distance approx. 0.5 m).
ð The second step of the calibration procedure starts.
10. Wait until the second part of the calibration procedure has been completed.
ð The main screen is displayed again.
11. Remove the calibration rod from the weighing platform.
ð The device is ready to measure.
Device mode
Basic
Advanced
Expert
Service
You can specify whether the device calculates Body Mass Index (
Surface Area (
BSA
) as soon as the patient's weight and height are available.
You can also set the device so that it does not display the
culated, but waist circumference
NOTE
If the
Waist circumf.
tion is not possible.
Function available
can be entered manually.
Waist circumf.
setting is selected, automatic BMI/BSA calcula-
English
) or Body
BMI
BMI
or
BSA
value cal-
Configuration • 167