Herunterladen Diese Seite drucken

Seca 655 Bedienungsanleitung Seite 138

Vorschau ausblenden Andere Handbücher für 655:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

5.4
Connecting a barcode scanner (optional)
138 • seca 655/654; seca 455; seca 257/256/254; seca 453
3. Align the device by turning the foot screws.
Foot screw
ð The device is positioned horizontally when the air bubble of the spirit level
is in the precise center of the circle.
CAUTION!
Injury from a lack of stability
If the foot screws are screwed out too far, they may come loose from
the device. The device will then be unstable.
► Screw the foot screws out a maximum of 10 mm.
► If the device cannot be aligned horizontally with the screws
screwed out as far as possible, the setup location is unsuitable. Set
the device up in a suitable location.
Knurled screw
4. Tighten the knurled wheels in the direction of the arrow.
ð The foot screws are secured against being adjusted.
5. If an ultrasonic measuring rod is fitted, calibrate it
measuring
rod, page 166.
A barcode scanner can be connected to the USB interface of the multifunctional
display.
The barcode scanner is required for the following functions:
: Define network data in the
Configuration
transmit them to the device using the QR code:
tions, page 173
: Record patient and user IDs for transmitting measured results to
Operation
the
seca analytics 125
measurement, page 158
WARNING!
Injury
► Route the connector cable so that patients cannot become caught
or strangle themselves in it.
► Route the connecting cable so as to prevent patients and users
tripping.
Spirit
level
seca connect 103
è Setting up network func-
software or to an EMR System:
Air bubble
è Calibrating ultrasonic
software and
è Completing the

Werbung

loading

Diese Anleitung auch für:

654455257256254453