Reset To Default Configuration; Unlocking The U/I/R Operation Mode; Locking The Device Configuration - Elektro-Automatik PSI 8000 T Betriebsanleitung

Dc laboratory power supply
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Using the power supply
Notifications of the set/actual comparison
Example: The step from a lower set value to a higher set value
was not performed within the settling time
error is then notified as alarm, warning or signal.
U
or
U
Depending on the configuration of
I
is notified.
Example: The step from a higher set value to a lower set value
was not performed within the settling time
error is then notified as alarm, warning or signal.
U
or
U
Depending on the configuration of
I
is notified.
Part 2: The menu
Options
Options +
The menu entry
Options
tion:
Reset configuration
Enable R mode
Setup lock
7.7

Reset to default configuration

You can reset all modifications of the setup to the default setup
(the state the unit had when it was delivered).
After selecting the corresponding menu entry you will be prompted
again to submit the choice to reset your current, personal con-
figuration.
Attention!
Even if the device configuration has been locked by a PIN
it will be unlocked and overwritten!
Operating Guide
PSI 8000 T Series
Tsr. The supervision
or
U
Step response
the error
Tsf. The supervision
or
U
Step response
the error
leads you to following menu selec-
Reset configuration +
Are you sure ?
= YES
All modifications of the default setup are reset
= NO
No change
7.8

Unlocking the U/I/R operation mode

The U/I/R operation mode can only be used after it was unlocked
with a PIN code in menu
Internal resistance"):
Enable R mode +
Activate R mode via pin code: 0 0 0 0
Use here the PIN code you received from your dealer. Once
unlocked, the status can be verified with:
R mode available:
YES
The U/I/R operation mode is unlocked and can
be used
NO
U/I/R mode not enabled yet
This mode also has to be configured in the profile (see also „7.1
Defining operation parameters"). The resistance set value is then
adjustable from 0Ω up to 20* Unom/Inom (of the device).
7.9

Locking the device configuration

Setup lock +
It can be necessary, for security reasons, to lock the device con-
figuration from access. You can enter a PIN code here, consisting
of 4 numbers, each from 0 to 15.
Lock setup via
Enter the PIN code
pin code:
{0..15} {0..15} {0..15} {0..15}
The lock can only be disabled with the same PIN code or by
resetting the configuration with
latter one deletes the custom setup and should only be used is
case the PIN code has been forgotten.
Attention!
This only affects the user profile of the device, not the set
values or the rotary knobs on the front panel!