Configuring The Control Panel - Elektro-Automatik PSI 8000 T Betriebsanleitung

Dc laboratory power supply
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Using the power supply
Limits of the set value of current
I adj
= {I
} {I
}
adj.min
adj.max
Whereas I
= {0...I
} and I
adj.min
adj.max
You can define the lower and upper limit of the adjustable current
here. Set values which exceed these limits are not accepted,
whether from the control panel nor from the remote control via a
PC (communication with interface cards).
Limit of the set value of power (only models from 1 kW)
P adj max
= {0 kW... P
}
nom
You can define the upper limit of the maximum adjustable power
here. Set values which exceed these limits are not accepted,
whether from the control panel nor from the remote control via a
PC (communication with interface cards).
Limit of the set value of internal resistance
(Optional, only accessible with unlocked U/I/R mode)
R adj max
= {0Ω...20 * Ri
}
nom
If the U/I/R mode has been unlocked, you can set the upper limit
of the maximum adjustable internal resistance. Set values which
exceed these limits are not accepted, whether from the control
panel nor from the remote control via a PC (communication with
interface cards).
Operating Guide
PSI 8000 T Series
Default:
0 A, I
nom
= {I
...I
}
adj.max
adj.min
nom
Default:
P
nom
Default:
0Ω
7.4

Configuring the control panel

Control panel +
The menu page
Control panel
eters that are related to the graphical display and the control panel.
Configure how set values are adjusted
Accept set value
= direct
The set values are directly set to the power
stage, when changed with the rotary knobs
= return key
The changed set values are only set if
submitted with the
= from preset list
You can choose sets from the
List
with the rotary encoders and submit
them with the
Control panel lock
The control panel lock is only configured here.
Key lock
= except OFF
The control panel (buttons and rotary
knobs) will be locked, except for the
button
= enable
The control panel will be completely locked
= disable
No lock
The control panel lock is used to prevent from unwanted changes
to the set values or to the settings.
Note
This setting is only temporary. It is reset
device is switched on again or returns from mains blackout.
Sounds
Key sound
= YES
A short beep signalises a button press
= NO
No signal if keys are pressed
Alarm sound
= YES
If an alarm or warning occurs an acoustic signal is
emitted (beep) in short intervals
= NO
No acoustic signal for alarms/warnings
Date: 02-05-2016
EN
lets you configure all param-
Default:
direct
RETURN
button.
Preset
RETURN
button
Default:
except OFF
OFF
(=
disable) after the
Default:
NO
Default:
YES
53