Using the power supply

6.15.6 Display during the function run

Also see the overview on the previous page.
70.00 V
35.00 A
1400kW
Display of the actual values
On the left side of the display the actual values are shown in small
font. The status of the active control (CV/CC/CP) is displayed to
the right of the corresponding value.
20.00 V
15.00 A
1500kW
The set values of the sequence point, which will be
reached after the sequence has been processed, are shown on
the right side of the display
1
2
2/5
Status display of the function run. The remaining
repetitions of the function (1) and of the sequence (2), as well as
the current sequence (2/_) and the momentarily active sequence
point (_/5) are displayed.
Function manager is halted or wasn't started yet
Function manager is running
15:05 m
The elapsed time since the function gene-
rator was started is also displayed. The time display is stopped
when the function manager stops. The STEP, RUN or GO keys
are used to run the function manager in several ways. The time
display will then continue to count.
{ON,OFF}
State of the power output
Besides the state of the power output an alarm, a warning or a
signal can be displayed.

6.15.7 Controlling the function manager

The interactive control panel provides keys to control the function
manager. You can halt, continue, reset it to the starting point or
exit the function by using these keys.
50
Date: 02-05-2016
Before the function manager is really setting the power supply
you can simulate the function on the display. During this
- the output is not switched on and
- the sequence points are processed step by step and can be
verified this way.
The execution is also controllable via communication with an
interface card. Here you can additionally set one stop point at
one of the 50 sequence points. This sequence point is processed
and the sequence/function is then halted.
The ESC button exits the function manager and returns
to the former state of the power supply.
The STEP button is used to run a sequence stepwise. The
current sequence point is executed after the button was pressed.
After the „step" has been executed, the set values, which are
displayed in upper right corner of the display, are set.
The RUN button starts the function manager and the
function is run as it was defined. The sequence points are then
processed consecutively.
Example for a simulation during standby:
Use the GO button to continue the function after it was
stopped.
Alternatively, you can reset the function manager to the
start of the current function with the NEW button.