6. USB-ScanSoft software
Function
Start conditions
Stop signal
EN
Limit value
Stop conditions
Stop signal
Limit value
26
Impact
Under the different start conditions, the desired condition must be
selected. With the setting "immediately", the measured value logging
begins immediately after the start.
For the other conditions, one or more threshold values must be entered
accordingly.
If these are under- or exceeded, the condition is fulfilled or not,
depending on the selected case.
The threshold values are entered in the free fields.
It is also possible to select which start signal should lead to the start of
the measurement. However, this is only possible if the cycle time is ≥ 5
ms.
Here, in analogy to the start conditions, the stop conditions are defined.
In addition to manual stop and measurement stop after a set time, stop
by certain threshold values is also available.
If these are under- or exceeded, the condition is fulfilled, depending on
the selected case.
The threshold values are entered in the free fields.
It is also possible to select which stop signal should lead to the stop of
the measurement. However, this is only possible if the cycle time is ≥ 5
ms.
WIKA operating instructions, models CPT2500, CPA2500 and USB-ScanSoft