Technical specifications
10.2 General technical specifications
Influence variables
Accompanying gases
Supply voltage
Table 10- 18 OXYMAT 7: Climatic conditions
Climatic conditions
Storage and transport
Max. permissible ambient temperature surrounding the
basic device during operation
Ambient humidity (rel. humidity) during storage, transport or
operation
2)
Restriction for installing an ULTRAMAT 7 analyzer module (combined operation): 5 to 45 °C
1)
Avoid condensation on the installed components!
2)
Table 10- 19 OXYMAT 7: Gas connections
Gas connections
With pipes
100
Deviation of the span gas
Analyzer module 24 V DC
(-5 %/+10 %)
Basic unit 85 V AC to 264 V AC/
47 Hz to 63 Hz
1)
≤ 1% of current span per 0.1 l/min change in flow or
50 vpm O
per 0.1 l/min change in flow within the permissi-
2
ble flow range
The larger value applies
Zero point deviation (cross-sensitivity) in accordance with
Table A.1 of EN 61207-3
< 0.1% of the current span
< 0.1% of the current span
- 30 to 70 °C
0 to 50 °C
< 90 %
Connecting socket for pipe with ∅ 6 mm
Compact Operating Instructions, 11/2019, A5E35134047-04