Technical specifications
10.2 General technical specifications
Electrical characteristics
during operation:
Electrical safety
Power unit cable
Table 10- 7
Wall housing: Purging gas pressure
Purging gas pressure
Continuous (recommended)
Continuous (maximum)
Transient (maximum)
Table 10- 8
Wall housing: Electrical inputs and outputs
Electrical inputs and outputs
Digital outputs (DO)
Digital inputs (DI)
Ethernet interface (bottom)
Service interface (inside at PU)
Table 10- 9
Wall housing: Option modules
Option modules
Option module 1.1
Option module 2.1
Option module 2.2
96
with one or two
in accordance with the standard requirements
CALOMAT 7 units
NAMUR NE21 (05/2006) and EN 61000-6-2:2005
in accordance with EN 61010-1, overvoltage category II
Max. Ø 10 mm
30 hPa above ambient
< 100 hPa above ambient pressure
165 hPa above ambient pressure
8, with changeover contacts, parameters can be freely assigned, e.g. for measuring
range identification;
max. load: 24 V AC/DC/1.7 A (total load for all 8 relay outputs in continuous operation
max. 160 W), isolated, non-sparking
8, designed for 24 V, isolated, freely parameterizable, e.g. for autoranging
Ethernet RJ 45, 100 MBit
Ethernet RJ 45, 100 MBit
12 digital outputs, with changeover contacts, max. load: 24 V AC/DC/1.7 A (total load
for all 12 relay outputs in continuous operation max. 244 W), isolated, non-sparking
8 digital inputs, designed for 24 V, isolated, freely parameterizable
6 analog outputs, 0/4 to 20 mA, designed for 24 V, isolated; maximum load 750 Ω
6 digital outputs, max. load: 24 V AC/DC/1.7 A (total load for all 6 relay outputs in
continuous operation max. 122 W), isolated, non-sparking
6 analog outputs 0/4 to 20 mA, designed for 24 V, isolated; max. load 750 Ω
4 analog inputs 0/4 to 20 mA, designed for 24 V, non-isolated, internal resistance
100 Ω
4 digital inputs, designed for 24 V, isolated
, max. 3 x 2.1 mm
2
2
Devices in explosion-proof design