Indicator fields for charging bays
"Charging progress" indicator
The "Charging progress" indicator is illuminated in
green as soon as a battery is inserted into the corre-
sponding charging bay. It indicates the charging status
of the battery. At full illumination, the charging process
is complete and the battery can be removed from the
charging bay.
"Action request" indicator
The "Action request" indicator is illuminated in orange
when a charging process cannot be properly carried out.
The "Charging progress" indicator goes out simultane-
ously.
Possible causes: bad contacts or excessive battery tem-
perature during charging. see Troubleshooting list
"Battery change recommended" indicator
The "Battery change recommended" indicator is illumi-
nated in orange in addition to the "Charging progress"
indicator when the charging process is complete. The
corresponding battery has failed the automatic battery
check. This battery can nevertheless still be used. How-
ever, it must be taken into account that an intraopera-
tive battery change may be necessary due to the
reduced performance.
7.

Validated reprocessing procedure

7.1

General safety notes

Note
Adhere to national statutory regulations, national and international stan-
dards and directives, and local, clinical hygiene instructions for reprocess-
ing.
Note
For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD, or possi-
ble variants of CJD, observe the relevant national regulations concerning
the reprocessing of products.
Note
It should be noted that successful reprocessing of this medical device can
only be guaranteed following prior validation of the reprocessing method.
The operator/reprocessing technician is responsible for this.
The recommended chemistry was used for validation.
7.5

Wipe disinfection for electrical devices without sterilization

Phase
Step
I
Wipe disinfection
RT:
Room temperature
Phase I
Remove any visible residues with a disposable disinfectant wipe.
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
Observe the specified application time (1 min minimum).
Clean the contacts in the charging bay with isopropanol or ethyl alco-
hol and a cotton swab. Do not use corrosive chemicals.
6
T
t
Conc.
Water quality
[°C/°F]
[min]
[%]
≥1
RT
-
-
7.2

General information

Only process chemicals that have been tested and approved (e.g. VAH or
FDA approval or CE mark) and which are compatible with the product's
materials according to the chemical manufacturers' recommendations
may be used for processing the product. All the chemical manufacturer's
application specifications must be strictly observed. Failure to do so can
result in the following problems:
Visual material changes
Material damage (such as corrosion, cracks, breaks, premature aging or
swelling).
7.3

Preparation before cleaning

The battery must be removed from the respective charging bay.
Unplug the power cord from the mains power socket.
Unplug power cord from the device connector 7.
7.4

Cleaning/disinfection

Product-specific safety instructions for the reprocessing procedure
Risk of electric shock and fire hazard!
Unplug the power cord before cleaning.
DANGER
Do not use flammable or explosive cleaning or
disinfecting solutions.
Ensure that no fluids will penetrate the product.
Damage to, or destruction of the product caused by
mechanical cleaning/disinfection!
CAUTION
Only clean and disinfect the product manually.
Do not sterilize the product under any circum-
stances.
Risk of damage to the product due to inappropriate
cleaning/disinfecting agents!
Only use cleaning/disinfecting agents approved
CAUTION
for surface cleaning. Follow the manufacturer's
instructions for the respective cleaning/disin-
fecting agent.
Do not introduce any fluid into the charging
bays.
Chemicals
Meliseptol Wipes sensitive