BTL7-A/E501-M_ _ _ _-J-DEXC-TA12
Magnetostrictive Linear Position Sensor – Rod Style
3
Construction and function (continued)
3.2

Function

The BTL contains the waveguide which is protected by an
outer stainless steel tube (rod). A magnet is moved along
the waveguide. This magnet is connected to the system
part whose position is to be determined.
The magnet defines the position to be measured on the
waveguide.
An internally generated INIT pulse interacts with the
magnetic field of the magnet to generate a torsional wave
in the waveguide which propagates at ultrasonic velocity.

Component overview

1)
Float
Pressure housing
2)
Magnet
O-ring flange (not
visible)
3)
Adapter
1) For use in liquid-level applications,
not included in scope of delivery (see Section 12.1 Floats)
2) Not included in scope of delivery (see Section 12.2 Magnets)
3) 1/2"-14 NPT to M20, optional (see Section 12.5 Conduit adapter)
4) Modified 1/2"-14 NPT according to FM 3615, 3.3.3, D, 1
5) M6x45 A4 hexagon socket screws (6x, included)
(replacement screw kit: BTL7-A-FK01-E-J-DEX).
6) Wiring information (see Section 4.5 Electrical connection)
7) Optional (see Section 12.4 Calibration device)
10
english
External housing GND
Housing cover
O-ring
4)
Conduit entry
Internal housing GND
Location of
order code
for replacement
electronics module 
Type of protection "db" and "t"
The component of the torsional wave which arrives at the
end of the waveguide is absorbed in a damping zone to
prevent reflection. The component of the torsional wave
which arrives at the beginning of the waveguide is
converted by a coil into an electrical signal. The travel time
of the wave is used to calculate the position. Depending
on the version, this information is made available as a
voltage or current with rising or falling gradient.
The following functions can be selected for the output
values:
Position
Velocity (with or without leading sign)
Differential position
Two outputs that can be independently assigned are
available. Two magnets can be used.
The entire function scope can only be
configured with the PC software "Micropulse
Configuration Tool". To do this, the USB
communication box must be connected (see
Accessories on page 29). If you do not have
the PC software, the BTL can be adjusted using
the calibration device (see from page 29).
Cover secondary retaining
screw
5)
Housing screws
Electronics module (including measuring range)
6)
Connection terminal
Calibration device
Flameproof enclosure
Retaining screws
Housing cover
7)