BTL7-A/E501-M_ _ _ _-J-DEXC-TA12
Magnetostrictive Linear Position Sensor – Rod Style
8

Teach-in with the calibration device

NOTICE
Interference in function
Teach-in while the system is running may result in
malfunctions.
► Stop the system before performing teach-in.
Initial situation:
BTL with magnet in measuring range
Activate teach-in
Œ Œ
► Hold down
for at least 4 s.
Set null point
► Bring magnet to the new null point.
Œ Œ
► Hold down
for at least 2 s.
⇒ The new null point is set once the button is released.
Set end point
► Bring magnet to the new end point.
‚ ‚
► Hold down
for at least 2 s.
⇒ The new end point is set once the button is released.
Invert curves
Œ Œ
‚ ‚
► Hold down
and
simultaneously for at least 4 s.
⇒ The curves of both outputs are inverted once the
buttons are released.
End teach-in
Œ Œ
‚ ‚
► Briefly press
and
simultaneously (< 1 s).
⇒ The current position value is displayed once the
buttons are released.
Any of the individual steps for settings can be
selected. The teach-in process can be ended at
any time.
1
= blue button
= gray button
2
www.balluff.com
Type of protection "db" and "t"
1
> 4 s
1
> 2 s
2
> 2 s
1
2
> 4 s
1
2
< 1 s
Flameproof enclosure
Displayed values (example)
At 0...10 V
At 4...20 mA
5.39 V
9.15 mA
5.39 V
9.15 mA
1.04 V
4.82 mA
0.00 V
4.00 mA
9.89 V
19.13 mA
10.00 V
20.00 mA