Beep signal
2 x short
3 x short
12.3 Troubleshooting
Event
Unexpected prosthesis
behaviour in various arm
positions, e.g. during over­
head use
Unexpected prosthesis
behaviour in case of differ­
ent weight settings, e.g.
when loading the prosthes­
is with excess weight
Hand cannot be opened
any more
Hand cannot be rotated any
more
-

12.4 Symbols Used

Type BF applied part
Declaration of conformity according to the applicable European directives
In some jurisdictions it is not permissible to dispose of these products with unsor­
ted household waste. Disposal that is not in accordance with the regulations of
your country may have a detrimental impact on health and the environment. Please
observe the instructions of your national authority pertaining to return and collec­
tion.
Manufacturer
Compliance with the requirements under "FCC Part 15" (USA)
Additional display
LED on charging
receptacle lights up
briefly
-
Cause
Use in this arm posi­
tion was not taken into
account during calib­
ration, or was incor­
rectly configured in the
app.
Faulty control
General fault
Event
Charging ends (charging plug disconnected
from charging receptacle)
Switching on the product
Battery voltage too low, product shuts off auto­
matically
Required action
Note arm position during calibration
Take the weight into account during
calibration
Adjust the values in the app under
"Settings > Advanced Settings"
Use mechanical overload protection:
Pull fingers open with great force
Rotate hand with great force
Switch the prosthesis off and back on
Consult O&P professional