11.2 Troubleshooting
Event
Lack of gripping force
Excessive friction in pronation and supination
direction
Insufficient friction in pronation and supination
direction
Hand operates too slowly in open direction
Hand operates too slowly in close direction
Finger cover damaged

1 Avant-propos

INFORMATION
Date de la dernière mise à jour : 2017-01-18
Veuillez lire attentivement l'intégralité de ce document avant d'utiliser le produit.
Cause/required action
Insufficient contact between hand and switch
unit.
Firmly screw the wrist nut onto the outer O-
ring.
Excessive frictional forces on the O-rings.
Take out O-ring(s).
Insufficient frictional forces on the O-rings, or
no O-rings installed.
Insert O-ring(s).
Second motor in operation because the pros­
thetic glove is too stiff.
Warm up the prosthetic glove (max.
60°C/140°F), then open the hand wide.
Allow to cool in open position.
Battery charge level low
Charge battery
Second motor in operation because the pros­
thetic glove is too stiff.
Close hand and warm prosthetic glove
from the outside (max. 60°C/140°F).
Battery charge level low
Charge battery
Cut open and pull off finger cover. Install new
finger cover.