Herunterladen Inhalt Inhalt Diese Seite drucken

Safety Shutoff; Beep Signals - Ottobock Myo Plus Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für Myo Plus:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 3
6.2 757L35 battery charger and 757B35=* rechargeable battery
6.2.1 Connect battery charger to the product
INFORMATION
The product cannot be used as long as the charging plug is connected to the charging recept­
acle. The product is deactivated for the duration of charging.

6.2.2 Safety Shutoff

The purpose of the prosthesis safety shutoff is to protect the battery; it is triggered in case of:
Excessively high or low temperature
Overvoltage or undervoltage
Short circuit
After a short circuit, the charging plug must be plugged into the charging receptacle and then
unplugged in order for the electronics to be activated.
6.2.3 Display of the current charge level
The battery charge level can be queried at any time.
1) With the prosthesis switched on, press the charging receptacle button and hold for less than
one second.
2) The LED display on the charging receptacle provides information on the current battery
charge level.
Charging
Event
receptacle
 
Battery fully charged (illuminated in green)
 
Battery 50% charged (illuminated in yellow)
 
Battery drained (illuminated in orange)

6.2.4 Beep signals

Beep signal
1 x long
2 x short
3 x short
26
1) Connect the charging plug to the charging receptacle on the pros­
thetic socket.
→ A correct connection between the battery charger and the pros­
thesis is indicated by feedback (see Page 32).
2) The charging process starts.
→ The prosthesis is switched off automatically.
3) Disconnect the prosthesis when the charging process is complete.
Additional display
-
LED on charging
receptacle lights up
briefly
-
Event
Turn off prosthesis on charging receptacle
Charging begins (charging plug connected
to charging receptacle)
Charging ends (charging plug disconnected
from charging receptacle)
Switch on prosthesis
Battery voltage too low, prosthesis shuts off auto­
matically