Phenomenon
Magnetic fields with
rated power frequen­
cies
Electrical fast transi­
ents/bursts
Surges
Line against line
Conducted interfer­
ence induced by high-
frequency fields
Voltage drops
Voltage interruptions
Interference resistance against wireless communication devices
Test fre­
Frequency
quency
band
[MHz]
[MHz]
385
380 to 390
450
430 to 470
710
704 to 787
745
780
810
800 to 960
870
930
EMC basic standard
or
test procedure
IEC 61000-4-8
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
Radio ser­
Modulation
vice
TETRA 400
Pulse modu­
GMRS 460,
FRS 460
± 5 kHz
deviation
1 kHz sine
LTE band 13,
Pulse modu­
17
GSM 800/90­
Pulse modu­
0,
TETRA 800,
iDEN 820,
CDMA 850,
GSM 800/90­
0,
LTE band 5
Interference immunity test level
30 A/m
50 Hz or 60 Hz
± 2 kV
100 kHz repetition rate
± 0.5 kV, ± 1 kV
3 V
0.15 MHz to 80 MHz
6 V in ISM and amateur frequency bands
between 0.15 MHz and 80 MHz
80% AM at 1 kHz
0% U
; 1/2 period
T
At 0, 45, 90, 135, 180, 225, 270 and 315
degrees
0% U
; 1 period
T
and
70% U
; 25/30 periods
T
Single phase: at 0 degrees
0% U
; 250/300 periods
T
Maximum
power [W]
1.8
lation
18 Hz
FM
1.8
0.2
lation
217 Hz
2
lation
18 Hz
Distance
Interfer­
[m]
ence
immunity
test level
[V/m]
0.3
27
0.3
28
0.3
9
0.3
28
35