The heel characteristics are adjusted with the function module. The following settings are pos­
sible:
Heel strike
Selected configuration (see fig. 2)
Soft
1
Moderate
2 (delivery condition)
Firm
3
► Turn the function module to the desired configuration (see fig. 6).
5.1.5 Pulling over the protective sock
>
The prosthesis alignment is complete and the prosthetic foot settings are final.
1) Remove the footshell from the prosthetic foot.
2) Pull the protective sock over the prosthetic foot such that there are no creases around the
sole.
3) Pull the footshell over the prosthetic foot again.
4) Tuck away the excess protective sock material around the foot adapter.
5.2 Preparing and connecting the foam cover
The connection cap serves as a detachable connection between the cosmetic foam cover and
footshell, and snaps precisely into the proximal brim of the footshell.
Required materials and tools: 634A58 Isopropyl Alcohol, 636N9 Contact Adhesive or
>
636W17 Plastic Adhesive
1) Cut the foam cover to length according to the patient's measurements, allowing for distal
compression of about 1 cm for transtibial and of about 4 cm for transfemoral covers.
2) Pull the foam cover over the prosthesis.
3) Join the connection cap to the footshell and mount the prosthetic foot.
4) Mark the outer contours of the connection cap on the distal face of the foam cover blank.
5) Dismount the prosthetic foot and remove the connection cap from the footshell.
6) Clean the connection cap with isopropyl alcohol.
7) Glue the connection cap to the face of the foam cover as marked and allow the glued joint to
set (approx. 10 minutes).
8) Mount the prosthetic foot and adapt the exterior cosmetic cover shape. Take into account
compression due to cosmetic stockings or SuperSkin.
6 Maintenance
► The prosthetic components should be inspected after the first 30 days of use.
► Inspect the entire prosthesis for wear during normal consultations.
► Conduct annual safety inspections.
7 Disposal
This product may not be disposed of with regular domestic waste in all jurisdictions. Disposal that
is not in accordance with the regulations of the country where the product is used may have a det­
rimental impact on health and the environment. Please observe the information provided by the
responsible authorities in the country of use regarding return, collection and disposal procedures.

8 Legal Information

All legal conditions are subject to the respective national laws of the country of use and may vary
accordingly.
8.1 Liability
The manufacturer will only assume liability if the product is used in accordance with the descrip­
tions and instructions provided in this document. The manufacturer will not assume liability for
14 | Ottobock
Position of insert
Insert removed, opening posterior
Insert posterior
Insert anterior