3.2 General Safety Instructions

CAUTION
Exceeding the service life and reuse on another patient
Risk of injury due to loss of functionality as well as damage to the product
Ensure that the approved service life is not exceeded (see section "Service life").
Only use the product for a single patient.
CAUTION
Use under unallowable environmental conditions
Risk of injury due to damage to the product
Do not expose the product to unallowable environmental conditions (see the section "Envir­
onmental conditions").
If the product has been exposed to unallowable environmental conditions, check it for dam­
age.
If damage is apparent or in case of doubt, do not continue using the product.
Take suitable measures if required (e.g. cleaning, repair, replacement, inspection by the
manufacturer or a specialist workshop etc.).
CAUTION
Excessive strain on the product
Fall due to breakage of load-bearing components
Use the prosthesis components according to the MOBIS classification (see section "Area of
Application").
Inform the patient.
CAUTION
Unallowable combination of prosthetic components
Risk of injury due to breakage or deformation of the product
Only combine the product with prosthetic components that are approved according to the
section "Combination possibilities".
Based on the instructions for use of the prosthetic components, verify that they may be com­
bined with each other.
NOTICE
Mechanical damage to the product
Change in or loss of functionality due to damage
Use caution when working with the product.
If the product is damaged, check it for proper function and readiness for use.
In case of changes in or loss of functionality, do not continue using the product (see "Signs
of changes in or loss of functionality during use" in this section).
Take any necessary measures (e.g. repair, replacement, inspection by the manufacturer's
customer service, etc.).
NOTICE
Mechanical overload
Impaired functionality due to mechanical damage
Check the product for damage prior to each use.