1.1 Operation and Functions

The VACUSON Group can be used for the suction of fluids and secretions. The pump power can be
continuously regulated and monitored through the Manometer.
The Vacuson can be used in the following areas:
- Chemistry
- Laboratory
- Medicine
- Intensive care

1.2 Technical Data

Voltage: ...................................................................... 115 V~ / 60 Hz or 230 V~ / 50 Hz (see unit label)
Current consumption: ................................................................................ Vacuson 40 115 VAC:.... 2 A
Fuses: ....................................................... 2 x T4A (for 115 V~ / 60 Hz) or 2 x T2A (for 230 V~ / 50 Hz)
Dimensions (W x H x D): ......................................................................................... 290 x 280 x 360 mm
Weight: .......................................................................................................................... 10 kg resp. 13 kg
Accuracy Limit Manometer:................................................................................................................. 5%
Suction Pump Capacity: ...................................................................................... 40 l/min resp. 60 l/ min
Suction Motor Power: ............................................................................................. 185 VA resp. 250 VA

1.3 Operating Environment

Relative Humidity:...................... max. 80 %
Temperature: .............................. 5 to 35 °C
Barometric Pressure:......... 800 to1060 hPa

1.5 Icons

: Attention accompanying documents
: Application of part Type BF
1275 : Conforms to EU standards
: Old electrical and electronic equipment must be disposed separately and may not be
included in regular domestic waste.
2

1 Description

- Dental laboratory
- Surgery
- Postoperative care

1.4 Transport and Storage Environment

Relative Humidity:.............................. max. 90 %
Temperature: ......................................5 to 50 °C
Barometric Pressure: ................ 700 to1060 hPa
NOUVAG AG
230 VAC:. 0.9 A
Vacuson 60
115 VAC:. 2.5 A
230 VAC:. 1.4 A