11 Technical Data
200 kHz LCR-Bridge R&S®HM8118
All data valid at 23 °C after 30 minutes warm-up.
Conditions
Test signal voltage
1 V
Open and short corrections performed
Measurement time
SLOW
Display
Measurement modes
Auto, L-Q, L-R, C-D, C-R, R-Q, Z-Θ, Y-Θ, R-X,
G-B, N-Θ, M
Equivalent circuits
auto, series or parallel
Parameters displayed
Value, deviation or % deviation
Averaging
2 to 99 measurements
Accuracy
Primary parameters
Basic accuracy
(Test voltage: 1.0 V,
measurement SLOW/MEDIUM,
autoranging mode, constant voltage OFF,
bias off). For FAST mode double the basic
accuracy values
Impedance: 100 MΩ
0.2% + I Z I / 1.5 GΩ
4 MΩ
1 MΩ
0.05% +
I Z I / 2 GΩ
25 kΩ
100 Ω
0.1% + 1 mΩ / I Z I
2.5 Ω
0.3% + 1 mΩ / I Z I
0,01 mΩ
20 Hz
Secondary parameters
Basic accuracy D, Q
±0,0001 if f = 1 kHz
Phase angle
±0,005° if f = 1 kHz
Ranges
|Z|, R, X
0,01 mΩ to 100 MΩ
|Y|, G, B
10 nS to 1.000 S
C
0,01 pF to 100 mF
L
10 nH to 100 kH
D
0,0001 to 9,9999
Q
0,1 to 9.999,9
θ
-180° to +180 °
-999,99 to 999,99 %
M
1 µH to 100 H
N
0,95 to 500
Measurement conditions and functions
Test frequency
20 Hz to 200 kHz (69 steps)
Frequency accuracy
±100 ppm
AC test signal level
50 mV
Resolution
10 mV
72
0.5% +
I Z I / 100 MΩ
0.1% +
I Z I / 1,5 GΩ
0.2% +
0.5% +
I Z I / 100 MΩ
5 mΩ / I Z I
+
0.2% +
I Z I / 10 MΩ
2 mΩ / I Z I
0.5% +
2 mΩ / I Z I
1 kHz
10 kHz
100 kHz
to 1.5 V
rms
rms
rms
Drive level accuracy
±(5 % + 5 mV)
Internal bias voltage
0 V
Resolution
10 mV
External bias voltage
0 V
Internal bias current
0 mA to +200 mA
Resolution
1 mA
Range selection
Auto and Hold
Trigger
Continuous, manual or external via interface,
binning interface or trigger input
Trigger delay time
0 ms to 999 ms in 1 ms steps
Measurement time (f ≥1 kHz)
FAST
70 ms
MEDIUM
125 ms
SLOW
0.7 s
Miscellaneous
Test signal level monitor
Voltage, current
Error correction
Open, short, load
Save/Recall
9 instrument settings
Front-end protection
V
max
(1 Joule of stored energy)
Low potential and
low current guarding
Ground, driven guard or auto (fused)
Constant voltage mode (25 Ω source)
Temperature effects R, L or C ±5 ppm/°C
Interface
Dual interface USB/RS-232 (R&S®HO820),
optional R&S®HO880 IEEE-488 (GPIB)
Safety
Safety class I (EN61010-1)
Power supply
110 V to 230 V ±10 %, 50 to 60 Hz, CAT II
Power consumption
approx. 20 W
Operating temperature
+5 °C to +40 °C
Storage temperature
-20 °C to +70 °C
Rel. humidity
5% to 80 % (non condensing)
Dimensions (W x H x D)
285 x 75 x 365 mm
Weight
approx. 4 kg
Accessories supplied: Line cord, operating manual,
R&S®HZ184 4-terminal kelvin test cable, R&S®HZ188
4-terminal SMD component test fixture
Recommended accessories:
R&S®HO118
Binning interface
R&S®HO880
Interface IEEE-488 (GPIB),
galvanically isolated
R&S®HZ42
19" rackmount kit 2RU
R&S®HZ72
GPIB-cable 2 m
R&S®HZ181
4-terminal test fixture including shorting
plate
R&S®HZ186
4-terminal transformer test cable
to +5,00 V
DC
DC
to +40 V
(fused 0.5 A)
DC
DC
<
2/C @ V
<200 V, C in Farads
max