Herunterladen Inhalt Inhalt Diese Seite drucken

R&S®Ho118 Description; Sorting Bin Preferences (Bins) - Hameg R&S HM8118 Benutzerhandbuch

Lcr-messgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Fig. 8.16: R&S®HO118 internal circuitry
The circuit is „active low", i.e.the voltage drops to 0V as
soon as the criterion for the respective BIN (set in the ins-
trument) is met. The function of the binning interface can
be tested by connecting a simple, passive component (e.g.
1kOhm resistance) to the instrument. In the BIN menu, set
a generous pass/fail criterion and measure the voltage of
PIN 25 (BIN2 on the circuit, BIN1 in the instrument) for the
25-pole plug to PIN 1 (GND). In case of „Pass", the voltage
should be 0V, in the case of „Fail", it should correspond to
the external voltage which must be applied to PIN 9.
For detailed information on the binning interface in the
context of the PIN and jumper assignment, refer to the
R&S®HO118 manual.
8.5.2 R&S®HO118 Description
BIN Type
Description
0...5
Pass BIN
This sorting bin is used if the measured value is
within the user-defined bin limit. If the measured
value is within this limit, it will be assigned to bin 0
(BIN 0). Outside the limit that is defined for bin 0,
the assignment within the limit for bin 1 (BIN 1) is
performed. This process is repeated until the limit
for bin 5 (BIN 5) is exceeded. If the measured value
exceeds the defined range limits for bins 1 to 5, it
will be assigned to the General-Failure bin.
6
Secondary
This sorting bin is used if the primary value is
Parameter
within the range for the sorting bins 0 ... 5 and
Failure BIN
only the secondary parameters exceed the limit for
sorting bin 6.
7
General Fai-
This sorting bin output is activated if the sorting
lure BIN
does not apply to one of the first 7 bins.
The Store/Recall feature enables you to determine a maxi-
mum of 9 binning configurations. Binning configurations
can also be operated via remote control interface. The
bridge R&S®HM8118 can sort components in up to 8 sepa-
rate bins: six pass sorting bins, one secondary parameter
sorting bin and one general sorting bin for errors. At any
given time, only one sorting bin (BIN) is activated.
Output signal:
Negative TRUE, open collector, opto-isolated, selectable
pull-ups.
Measurement types:
Since the R&S®HM8118 is used for classification, the num-
ber of measurement types is limited to the modes needed
for the characterization of components.
❙ R-Q: Resistance value and quality
❙ C-D: Capacitance value and loss angle
❙ L-Q: Inductance and quality
Sorting bins (BINs):
❙ Pass bin: Bin 0...5 for primary parameters
❙ Fail bin: Bin 6 for secondary parameters, bin 7 for general
errors (General Failure BIN).
❙ Maximum current for an output voltage of 1 V is 15 mA.
Index:
Analog measurement completed.
Measurement completed:
Complete measurement completed.
Alarm:
Notification about a known bug.
External trigger:
Opto-isolated, selectable pull-up, pulse width >10µs.
8.5.3

Sorting Bin Preferences (BINs)

The R&S®HM8118 must be in manual mode. Select the re-
spective function of the parameter to be sorted. As men-
tioned in the section „Measurement Types", all functions
can be used. To be able to enter binning parameters, press
the MENU key and select the option BIN. A binning inter-
fact must be integrated to be able to access the binning
menu.
Example:
Binning:
BIN Number:
BIN:
Nominal:
Low limit:
High limit:
Binning ON/OFF:
❙ ON: Binning function activated
❙ OFF: Binning function deactivated
BIN Number:
❙ Selecting the BIN number
❙ Bins 0 to 5 correspond to the primary pass bins
❙ Bin 6 corresponds to the secondary parameter failure bin
❙ Bin 7 (General Failure BIN 7) does not have a menu entry.
BIN OPEN or CLOSED:
❙ OPEN: The respective BIN is activated.
❙ CLOSED: The respective BIN is deactivated.
❙ At least the first bin must be activated.
Measuring Equipment
ON
0
Open
100.0
-4.0%
+5.0%
65

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis