4. Are the ends of the actuator stirrup flush with the functional knee part?
Event
Noise development
5. Is the position of the actuator stirrup correct?
Event
Collision between
actuator stirrup and
joint frame
6. Can the feedback signals be heard clearly enough?
Event
Feedback signals
cannot be heard
clearly enough
6.10 Finishing the functional cosmesis
Complete the following steps to finish the functional cosmesis:
1) Assemble the functional shank, knee joint and prosthetic foot.
27
Cause
Incorrect position of the actu­
ator stirrup ends
Cause
Incorrect position of the actu­
ator stirrup
Cause
Improper volume setting on
the joint
Correction
1) Loosen a screw of the
actuator
functional knee part.
2) Reposition
stirrup (see fig. 12).
3) Tighten the screws of the
actuator stirrup.
4) Repeat steps 1-3 with the
second screw.
Correction
1) Remove
knee part from the joint.
2) Position the actuator stir­
rup (see fig. 10).
3) Attach the functional knee
part to the joint.
Correction
► Adjust the volume of the
joint
feedback
(note the instructions for
use:
3B1-2/3B1-2=ST
"647G1148",
3B1-3/3B1-3=ST
"647G1380" and C-Leg
3C98-3/3C88-3
"647G1375").
2) Fold over the coated ankle part of the func­
tional stocking (coating faces out).
stirrup
on
the
the
actuator
the
functional
signals
Genium
Genium