W W a a r r n n i i n n g g / / C C a a u u t t i i o o n n / / N N o o t t e e D D e e f f i i n n i i t t i i o o n n
The words W W A A R R N N I I N N G G , C C A A U U T T I I O O N N , and N N O O T T E E carry special meanings and should be carefully reviewed.
W W A A R R N N I I N N G G
Alerts the reader about a situation which, if not avoided, could result in death or serious injury. It may also describe
potential serious adverse reactions and safety hazards.
C C A A U U T T I I O O N N
Alerts the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the
user or patient or damage to the product or other property. This includes special care necessary for the safe and effective
use of the device and the care necessary to avoid damage to a device that may occur as a result of use or misuse.
N N o o t t e e - - Provides special information to make maintenance easier or important instructions clearer.
S S u u m m m m a a r r y y o o f f s s a a f f e e t t y y p p r r e e c c a a u u t t i i o o n n s s
Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
W W A A R R N N I I N N G G
• Install the cot fastener by qualified personnel only. Improper installation could result in injury to the patient or operator.
• Always make sure that all cots meet the installation specifications for the Stryker Cot Fastener System. Injury may result
if a non-compatible cot is used in the Model 6385, 6386, 6387, 6388, or 6389 Cot Fastener System.
• Always test the cot fastener using BS:EN1789 as a minimum. Improper testing may result in injury.
• Always replace the cot fastener system if it has been involved in an accident to avoid the risk of injury due to product
damage.
• Install the floor plate by qualified personnel only. Improper installation could result in injury to the patient or operator.
• Do not use hand or fingers to press down on the release lever when the hooks are open. The fastener closes with a
strong spring action.
• Always wipe the product with clean water and dry after cleaning. Some cleaning products are corrosive in nature and
may cause damage to the product. Failure to properly rinse and dry the product leaves a corrosive residue on the
surface of the product and may cause premature corrosion of critical components.
C C A A U U T T I I O O N N
• Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this
manual.
• Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids its warranty.
• Do not use screws smaller than the 3/8 in. diameter specified to anchor the floor plates.
• Always make sure that the cot is in the lowest height position before you load the cot into the fastener. Press the retract
(-) button until the cot stops moving down to make sure that the cot will properly lock into the fastener.
• Do not press the extend (+) button while the cot is secured into the fastener to avoid injury and product damage.
EN
2
6385-109-005 Rev A.0