S S e e t t u u p p
W W A A R R N N I I N N G G
• Install the cot fastener by qualified personnel only. Improper installation could result in injury to the patient or operator.
• Always make sure that all cots meet the installation specifications for the Stryker Cot Fastener System. Injury may result
if a non-compatible cot is used in the Model 6385, 6386, 6387, 6388, or 6389 Cot Fastener System.
• Always test the cot fastener using BS:EN1789 as a minimum. Improper testing may result in injury.
• Always replace the cot fastener system if it has been involved in an accident to avoid the risk of injury due to product
damage.
The condition of the cot fastener system is the responsibility of the user. Have a qualified service person use the following
list and operation instructions to check functionality.
Floor plates are included and intact (Model 6385 only) ( Installing the floor plate (page 7))
Foot end fastener and head end fastener are included and intact ( Installing the head end and foot end fastener (Model
6385) (page 9), Installing the head end and foot end fastener (Model 6386 and 6387) (page 9), and Installing the head
end and foot end fastener (Model 6388 and 6389) (page 11))
Foot end fastener locks and unlocks properly ( Installing the head end and foot end fastener (Model 6385) (page 9),
Installing the head end and foot end fastener (Model 6386 and 6387) (page 9), and Installing the head end and foot end
fastener (Model 6388 and 6389) (page 11))
I I n n s s t t a a l l l l a a t t i i o o n n
I I n n s s t t a a l l l l i i n n g g t t h h e e f f l l o o o o r r p p l l a a t t e e
The Stryker Cot Fastener System for the floor plate mounting model is compatible only with cots that conform to the floor
plate mounting installation specifications listed below and comply with section 3.11.5 and 3.11.7 of the Federal
Specification for the Star-of-Life Ambulance, KKK-A-1822 Revision E.
W W A A R R N N I I N N G G
• Install the floor plate by qualified personnel only. Improper installation could result in injury to the patient or operator.
• Always make sure that all cots meet the installation specifications for the Stryker Cot Fastener System. Injury may result
if a non-compatible cot is used in the Model 6385, 6386, 6387, 6388, or 6389 Cot Fastener System.
C C A A U U T T I I O O N N - - Do not use screws smaller than the 3/8'' diameter specified to anchor the floor plates.
1. Determine the desired position of the cot inside the vehicle to allow for proper clearance around the cot.
N N o o t t e e - - Federal ambulance specifications regarding aisle width and attendant seating may affect positioning of the
mounting brackets. If specifications do not apply, the installer should be certain that there is adequate clearance for
proper patient access, traction splint projection, rear door closure, etc.
2. Using the dimensions shown in Figure 2, locate and drill the points where the floor plates will attach.
N N o o t t e e
• The centerline of the cot must be parallel to the vehicle wall.
• Make sure that you space the floor plates evenly around the centerline of the cot.
• Stryker recommends 370 mm from the centerline of the cot to the vehicle side wall to allow for wheel rotation.