Herunterladen Inhalt Inhalt Diese Seite drucken

Connector Pin Assignment - HBM t40 Montageanleitung

Torque flange drehmomentmessflansch
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

T40
usually be overcome by connecting the rotor directly to ground, for instance
by a wire loop. The stator should be fully grounded in the same way.

4.3 Connector pin assignment

The stator housing has two 7 pin device plugs (Binder 723) and one 8 pin
device plug.
The supply voltage connections and shunt signal connections of plugs 1 and 3
are each electrically interconnected, but are protected against compensating
currents by diodes. There is also an automatically resetting fuse (multifuse) to
protect the supply connections against overload by the stator.
Assignment for plug 1:
Supply voltage and frequency output signal.
Binder 723
6
1
5
7
2
3
4
Top view
1)
Complementary signals RS 422; for cable lengths of 10 m and longer, we recommend to
use a termination resistor R=120 ohms between wires (wh) and (rd).
CAUTION
Torque flanges are only intended for operation with a DC supply voltage.
They must not be connected to older HBM amplifiers with square wave
excitation. This could destroy the connection board resistances or
cause other faults in the amplifiers.
A2463 5.0 en/de
Binder
Assignment
pin
1
Torque measurement signal (frequency output;
1)
5 V
2
Supply voltage 0 V;
3
Supply voltage 18 V to 30 V
4
Torque measurement signal
(frequency output; 5 V
5
Meas. signal 0 V;
6
Shunt signal trigger 5 V to 30 V
7
Shunt signal 0 V;
Shielding connected to enclosure ground
1)
)
symmetrical
19
Color
Sub D
code
con
nector
pin
wh
13
bk
5
bu
6
rd
12
gy
8
gn
14
gy
8
HBM