Herunterladen Inhalt Inhalt Diese Seite drucken

Settings; Output Frequency Settings - HBM t40 Montageanleitung

Torque flange drehmomentmessflansch
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

T40
6

Settings

6.1 Output frequency settings

The standard frequency setting (DU2) is 60 kHz ± 30 kHz. The configurations
SU2 10 kHz ± 5 kHz and HU2 240 kHz ± 120 kHz can also be ordered.
The customer can also change the configuration by rearranging jumpers on
the stator electronics. To do this, loosen the Phillip's head screws in the stator
cover, remove the cover and move the jumper as shown in the illustration.
Then replace the cover and screw the screws back in. The setting only takes
effect the next time you start up, the change is not made during ongoing
operation.
NOTE
When rearranging the jumper, make sure that you do not lose it. It
cannot be replaced separately by HBM.
Fig.6.1:
Jumper settings
A2463 5.0 en/de
Jumper position for
D
D
D
D
output signal:
D
D
D
D
240 kHz $120 kHz
J4 J3 J2 J1
Jumper position for
D
D
D
output signal:
D
D
D
Jumper
10 kHz $5 kHz
J4 J3 J2 J1
Jumper position for
D
D
D
D
output signal:
D
D
D
D
60 kHz $30 kHz
J4 J3 J2 J1